Selected product option for Digital pressure probes for differential and absolute pressure for PRO-line. All information below applies to the selected product option.
Pressure checks on air and dry gas systems need a probe that is compact enough for handheld use and stable enough for demanding field work. The DX 210 supports differential pressure measurement across five ranges from ±2.5 hPa to ±700 kPa, giving users the right span for low-pressure checks as well as higher differential pressure tasks.
Temperature compensation over 0…+50 °C supports dependable readings, while factory calibration lets the probe go straight into use. Calibration data are stored in the probe, which supports interchangeability with PRO-Line handheld meters without recalibration.
The stainless-steel body, stainless-steel ports for Ø6x1 mm hoses and ingress protection (IP65) make the DX 210 a robust choice for portable measurement in air and non-aggressive dry gases. The digital M12 DX-Sensor interface connects to PRO D01, PRO D05, PRO 925 and PRO 595 meters.
DX 210 probes measure differential pressure, while DX 240 probes measure absolute pressure.
A 5-pin M12 connector links the probe to compatible PRO Line handheld meters with digital inputs.
Stainless-steel ports accept Ø6x1 mm hoses with an internal diameter of Ø4 mm.
The internal measuring rate can be set to Low or High in supported handheld meters.
The probes provide last value, minimum, average and maximum channels from the sensor.
The measurement is temperature compensated across 0 to +50 °C.
Three adjustment points and an offset function are available when User calibration type is selected.
The probes are intended for air and non-aggressive dry gases only.
| Type of media |
Air, Non-aggressive dry gases
|
|---|
| Sensor type pressure |
MEMS
|
|---|---|
| Measurement rate |
1000 /s
|
| Accuracy pressure |
Depending on sensor model
|
| Sensor protection |
IP65 (Keep measuring input dry for precise measurement)
|
| Pressure type |
differential
|
| Operating temperature |
-20...+85 °C
|
|---|---|
| Operating relative humidity |
0...95 %RH
|
| Condensation |
not permitted
|
| Storage temperature |
-20...+85 °C
|
| Weight |
130 g
|
|---|
| Power consumption |
Normal: 3.5 mA
Fast: 5.5 mA |
|---|
| Connection |
5-pole M12
|
|---|---|
| Probe Diameter |
21.7 mm
|
| Probe length |
62 mm
|
| Material |
Stainless Steel
|
| Process connection |
Tube connection 4 mm
|
|---|
| Certificate |
CE Declaration: Declaration of conformity required where applicable under EU legislation. Manufacturers declaration, nessasary for selling to european market. Requires -prodcut depending-, fulfillment of regulation, like RoHs, LVD, EMC, ATEX, product related |
|---|---|
| Standard / Compliance |
RoHS: Restriction of Hazardous Substances compliance. Manufacturers declaration of compliance to regulation, company/site related REACH: Compliance with REACH requirements for substances and SVHC information. Manufacturers declaration of compliance to regulation, company/site related |
| Compliance Requirement |
EMC directive: Electromagnetic Compatibility |