Selected product option for Waterproof conductivity meter. All information below applies to the selected product option.
Low-conductivity liquids need a meter that can resolve small changes without adding complexity to daily work. ECO 523 is designed for purest water measurement and service tasks in osmosis systems, filter technology, laboratories, pharmaceutical applications and quality assurance.
The stainless steel/PEEK 2-pole measuring cell supports high-resolution conductivity measurement up to 0.001 µS/cm. The device covers conductivity up to 100.0 µS/cm and specific resistance up to 20.00 MΩ/cm, with integrated temperature measurement and compensation modes for nLF, linear behavior and weak NaCl solutions according to EN 60746-3.
For work at the process edge, in the lab or during service visits, the handheld housing combines ingress protection (IP65/IP67), a break-proof ABS body and a backlit 3-line display. Hold, Min/Max and overhead display functions help users read and record values cleanly, even when access is awkward.
ECO 522 uses a 2-pole graphite measuring cell; ECO 523 uses a 2-pole stainless steel/PEEK measuring cell.
ECO 522 measures conductivity, salinity and TDS; ECO 523 measures conductivity, specific resistance and NaCl.
The meters offer deactivated or nLF compensation according to EN 27888; ECO 523 also includes LIN and NaCl modes.
Four long-lasting buttons operate the illuminated 3-line display, Hold function, Min/Max values and display rotation.
Temperature zero point and gradient correction, plus conductivity gradient correction, are available via the adjustment menu.
The devices operate with 2 x AA batteries and include a battery status indicator with depleted battery message.
| Number of analog inputs |
1
|
|---|---|
| Accuracy temperature |
±0.3 °C
|
| Resolution temperature |
0.1 °C
|
| Temperature range |
-5...+105 °C
|
| Measurement rate |
10 /s
|
| Conductivity range |
0...2 µS/cm
|
| Accuracy Conductivity |
Typ. ± 1 % of measured value ± 0.5 % FS
|
| Temperature compensation |
Off: deactivated
nLF: non-linear, acc. to EN 27888 LIN: linear with variable coefficients NaCI: For weak NaCl solutions in accordance with EN 60746-3 |
| Resolution |
0.1 µS/cm
|
| Operating temperature |
-20...+50 °C
the conductivity measuring cells can be exposed temporarily to temperatures of up to 100 °C and permanently to temperatures of up to 80 °C |
|---|---|
| Operating relative humidity |
0...95 %RH temporarily condensing
|
| Storage temperature |
-20...+70 °C
|
| Degree of protection case |
IP65, IP67
|
| Dimensions |
108 mm x 54 mm x 28 mm
|
|---|---|
| Materials |
ABS, Polyester, Case/housing: break-proof ABS, front panel: polyester
|
| Weight |
210 g
|
| Battery life |
1000 h
|
|---|---|
| Battery life description |
Service life > 1000 hours with alkaline batteries (without backlight)
|
| Battery type |
Battery 2x AA alkaline
|
| Power consumption |
2.2 mA, approx. 3.5 mA with backlight
|
| Cable |
1.2 m Fixed cable
|
|---|---|
| Probe Diameter |
12 mm
|
| Probe length |
75 mm
|
| Material |
Stainless Steel
|
| Wetted parts |
Stainless steel 1.4404/316L, Stainless Steel 1.4435, PEEK
|
|---|
| Display presence |
Yes
|
|---|---|
| Description |
3-line with battery status indicator, Backlight, protected by shatterproof screen, overhead display at the push of a button
|
| Number of controls |
4
|
| Type of controls |
4 Button
|
| Display type |
LCD backlit
|
| Certificate |
CE Declaration: Declaration of conformity required where applicable under EU legislation. Manufacturers declaration, nessasary for selling to european market. Requires -prodcut depending-, fulfillment of regulation, like RoHs, LVD, EMC, ATEX, product related |
|---|---|
| Standard / Compliance |
RoHS: Restriction of Hazardous Substances compliance. Manufacturers declaration of compliance to regulation, company/site related REACH: Compliance with REACH requirements for substances and SVHC information. Manufacturers declaration of compliance to regulation, company/site related |